Chiropractic X-Ray Rooms

Primarily utilizes an electrically charged cathode tube to produce ionizing x-ray gamma photons to still-image the spine and some soft tissue related to skeletal bone structures and alignment for diagnosis. Typical energy levels are 50kvp-125kvp. Rooms are typically smaller in size from 8'0" x 10'0" and up to 12'0" x 16'0". These rooms require shielded walls, an x-ray glass control window or x-ray control barrier with a x-ray view window minimum 8" x 10". Shielding must be calculated by a physicist or radiation expert retained by the facility. Shielding is based on the specific imaging equipment utilized and typically will involve several different lead thicknesses depending upon primary beam and secondary scatter radiation fields, workload, surrounding occupancy factors and other considerations.