RADIATION SHIELDING CONSIDERATIONS
Per the current NCRP Report #147(2004)1.5 General Concepts,states: "The term "qualified expert" used in this report is defined as a medical physicist or medical health physicist who is competent to design radiation shielding for medical x-ray imaging facilities.The qualified expert is a person who is certified by the American Board of Radiology,American Board of Health Physics,or the Canadian College of Physicists in Medicine.
Radiation shielding shall be designed by a qualified expert to ensure the required degree of protection is achieved.
The qualified expert should be consulted during the early planning stages since the shielding requirements may affect the choice of location of radiation facilities and type of building construction.The qualified expert should be provided with all pertinent information regarding the proposed radiation equipment and its use,type of building construction,and occupancy of nearby areas. It may be necessary to submit the final shielding drawings and specifications to pertinent regulatory agencies for review prior to construction".
LEAD SHIELDING REQUIREMENTS SHOULD BE CALCULATED AND DETERMINED BY A CERTIFIED
RADIATION HEALTH PHYSICIST BASED ON THE FOLLOWING FACTORS:
1) ENERGY: HOW POWERFUL IS THE X-RAY UNIT(S) OUTPUT IN KvP?
TYPICALLY, THE HIGHER THE OUTPUT, THE HIGHER THE SHIELDING REQUIREMENT.
2) WORKLOAD: HOW MANY EXPOSURES PER WEEK / MONTH / YEAR? MAXIMUM
PATIENT VOLUME GROWTH SHOULD BE PROJECTED IN, AS RADIATION IS CUMULATIVE AND OVEREXPOSURE
CAUSES BIOLOGICAL DAMAGE ON THE CELLULAR LEVEL. AS AN EXAMPLE, IN CALIFORNIA, IT IS
LIMITED TO A MAXIMUM PERMISSIBLE EXPOSURE LEVEL OF 100 mrem PER YEAR, IN ADDITION TO THE
360 mrem OF RADIATION THE AVERAGE PERSON IS EXPOSED TO PER YEAR DUE TO ENVIRONMENTAL
FACTORS (I.E.: ULTRA VIOLET, SOLAR, RADON GAS AND ATMOSPHERIC RADIATION).
ALWAYS CHECK WITH ALL LOCAL HEALTH AND SAFETY LAWS AND CODES FOR CURRENT COMPLIANCE
REQUIREMENTS.
3) ORIENTATION: EACH WALL SECTION WILL BE CALCULATED BY YOUR
PHYSICIST IN RELATION TO THE DIRECTION OF THE PRIMARY BEAM TARGET (WHERE IT WILL BE AIMED)
AND SCATTER (SECONDARY) RADIATION OF THE X-RAY UNIT, AS WELL AS FLOOR OR WALL BUCKY
CASSETTE HOLDER (TARGET AREA).
4) DISTANCE: RADIATION DISSIPATES AS DISTANCE INCREASES.
TYPICALLY, THE CLOSER A PARTITION IS TO THE RADIATION / X-RAY SOURCE, THE HIGHER THE LEAD
SHIELDING REQUIREMENT.
5) OCCUPANCY: ONE OF THE MOST IMPORTANT FACTORS OF CALCULATION IS
HOW MUCH TIME IN A DAY THAT A SURROUNDING ROOM COMMON TO THE X-RAY ROOM WILL BE OCCUPIED
BY YOUR PERSONNEL OR THE PUBLIC. (I.E.: A STORAGE ROOM WOULD HAVE A LOWER OCCUPANCY FACTOR
AND A WAITING ROOM OR OFFICE MAY BE HIGH). THEREFORE, IT IS CALCULATED ROOM BY ROOM FOR
ALL ROOMS CONNECTED (OR COMMON) TO THE X-RAY ROOM. IF THERE IS NO OCCUPANCY POTENTIAL,
THEN TYPICALLY NO SHIELDING WILL BE REQUIRED (I.E.: FLOOR OR ROOF OF A SINGLE STORY
BUILDING).
6) MATERIAL: SOMETIMES A PHYSICIST WILL TAKE INTO CONSIDERATION
THE EXISTING OR PROPOSED CONSTRUCTION MATERIAL OF THE WALL PARTITION OR FLOORING MATERIAL,
AS HEAVY DENSITY MATERIALS CAN ATTENUATE (SHIELD) RADIATION TO A CERTAIN DEGREE, SUCH AS
CONCRETE, STEEL, PLASTER, BLOCK OR MULTIPLE LAYERS OF DRYWALL. THIS MAY REDUCE, OR IN SOME
CASES, ELIMINATE YOUR LEAD SHIELDING REQUIREMENTS, DEPENDING ON THE VALUES OF THE PREVIOUS
FIVE FACTORS TO BE CONSIDERED.
- THERE MAY ALSO BE
OTHER CONSIDERATIONS, SUCH AS X-RAY FILM TO BE PROTECTED FROM RADIATION EXPOSURE TO
PREVENT "FOGGING" AND THE SHIELDING OF VARIOUS RADIATION SENSITIVE
EQUIPMENT, OR DESIGNATIONS OF PERIMETER OCCUPANCIES AS
"CONTROLLED" OR "UNCONTROLLED"
NOTES:
- ALL PENETRATIONS
WITHIN A SHIELDED WALL PARTITION MUST BE SHIELDED WITH EQUAL OR GREATER VALUE.
- MINIMUM HEIGHT OF
WALL SHIELDING IS 7'0" A.F.F. PER NCRP REPORT #49 (NATIONAL STANDARD)
- LEAD SHIELDING FOR
PARTITIONS MUST BE BONDED TO RIGID PANELS (I.E.: DRYWALL, PLYWOOD, OR PANELING) FOR
STRUCTURAL STRENGTH.
- SHIELDING
REQUIREMENTS MAY CHANGE IF ANY ONE OF THE FIVE BASIC FACTORS CHANGE.
- ALL RADIATION
PRODUCING SOURCES MUST BE LICENSED AND REGISTERED.
- TYPICAL DIAGNOSTIC
ROOMS: X-RAY, RADIOGRAPHIC, FLUOROSCOPIC, C.T. SCAN, CATH LAB, ANGIOGRAPHY, CAT SCAN,
MAMMOGRAPHY, ETC.
- RADIATION THERAPY
ROOM CALCULATIONS ARE MUCH MORE INVOLVED THAN DIAGNOSTIC DUE TO THE HIGH ENERGY LEVELS AND
OTHER CONSIDERATIONS TO FACTOR IN.
- END USER
SHOULD ALSO HAVE ALL SHIELDING PROJECTS COMPLETELY SURVEYED BY A QUALIFIED
RADIATION HEALTH PHYSICIST AFTER INSTALLATION AND PRIOR TO USE

"Over
65 Years of Manufacturing Excellence"

[ Wall Systems ] [ Shielded Doors ] [ Control Windows ] [ Other Products ] [ Shielding Information ] [ Specifications ] [ Contact Us ]

HOME PAGE
RAY-BAR ENGINEERING CORPORATION
697 W. Foothill Blvd
P.O. BOX 415
Azusa, CA 91702-0415 |
Toll
Free: 1(800) 567-XRAY (9729)
Telephone: (626) 969-1818
Fax: (626) 969-6510 |
© 2000-2008 All Rights Exclusive and Reserved Ray-Bar
Engineering Corporation
THESE PRODUCTS AND
MATERIALS ARE FOR MEDICAL AND INDUSTRIAL USE AND APPLICATIONS ONLY
(NOT TO BE UTILIZED FOR RESIDENTIAL APPLICATIONS)